Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: MAGNETOM Skyra magnetic resonance diagnostic devices Model # 10432915

Recall number
Z-2343-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
MAGNETOM Skyra magnetic resonance diagnostic devices Model # 10432915
Status
Terminated
Initiated
2019-07-09
Posted
not on file
Terminated
2020-04-23
Reason
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
Root cause
Device Design
Action
1. review the System Owner Manual chapter 1.3.3 Quench emergency plan and the section regarding defective vent line. 2. follow the standard emergency procedures you have in place in the event of system failures. 3.A Siemens Service Technician will conduct an inspection to verify that your system has a vertical outlet. If a vertical outlet is present, no further action is required. In the event your system is equipped with a horizontal outlet, the Siemens Service Technician will ensure that the port for the drop-off plate is adequately blanked off or sealed. 4.service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. 5. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. By signing this document you are acknowledging that you have read and understand the content therein. If you have any question, please contact Meredith Adams, Telephone 610-448-6461 and E-Mail: Meredith.adams@siemens-healthineers.com
Quantity
44
Distribution
AL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY
Lot, serial or model codes
Serial # 46066 45732 46182 45735 45805 46171 45693 45886 45988 45881 46144 46044 46176 46179 46032 46238 145161 46127 45774 45968 46196 46175 45681 46007 46021 46105 46142 45928 45815 46150 145302 45878 46239 46026 145403 45667 145536 45970 45839 45982 46145 46298 45777 46294
510(k) numbers
["K111242"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register