Spacelabs Healthcare Inc: 91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.

Recall number
Z-2343-2016
Recalling firm
Spacelabs Healthcare Inc
Product
91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP. Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP. Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP. Option I: SpO2, Temperature x2, & adult/neonate NIBP.
Status
Terminated
Initiated
2016-07-01
Posted
2016-08-04
Terminated
2017-05-22
Reason
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
Root cause
Vendor change control
Action
Spacelabs expanded the recall and sent an Urgent - Medical Device Correction letter dated August 30, 2016, to additional affected customers via priority mail (return receipt requested). On September 6, 2016, a customer letter (translated as necessary) was sent via email to all international subsidiaries and distributors of record. ********************************************************************************************** Spacelabs Healthcare Inc, sent an Urgent - Medical Device Correction letter dated July 1, 2016, to all affected customers (return receipt requested). On July 8, 2016, a customer letter (translated as necessary) was sent via email to all international subsidiaries and distributors of record. Spacelabs Healthcare will contact customers to schedule a convenient time for a Spacelabs service technician to update their affected Command Module(s) at no cost. For additional information or technical assistance, please contact: Technical Support 1-800-522-7025 and select (2) for Technical Support
Quantity
3801 modules and 116 PCBAs & valves distributed in the US; 1297 modules and 366 PCBAs & valves distributed outside the US
Distribution
Worldwide Distribution - USA Distribution and to the countries of : AFGHANISTAN, ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BRUNEI DARUSSALAM; CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DOMINICAN REPUBLIC; ECUADOR, EGYPT, ESTONIA, FRANCE, GERMANY, GUATEMALA, HONG KONG, INDIA, IRAN (ISLAMIC REPUBLIC OF), ISRAEL, ITALY, JORDAN, KUWAIT, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, PAKISTAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, Slovakia, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM.
Lot, serial or model codes
Manifold Lot Codes: 128280, 129420, 129940, or 155640.
510(k) numbers
["K103142"]
PMA numbers
not on file
Product code
DSI
Firm FEI
3010157426
Firm city
Snoqualmie
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register