FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: GE HealthCare OEC 9900 Elite C-arm Systems.

Recall number
Z-2342-2025
Recalling firm
GE Medical Systems, LLC
Product
GE HealthCare OEC 9900 Elite C-arm Systems.
Status
Open, Classified
Initiated
2025-07-02
Posted
2025-08-22
Terminated
not on file
Reason
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Root cause
Radiation Control for Health and Safety Act
Action
GE HealthCare will send a customer letter providing safety instructions as an immediate mitigation, to communicate the risk of additional radiation and that the equipment should only be used by personnel properly trained and authorized in the use of ionizing radiation. A user should observe proper safety practices during equipment operation as required by the operator manual. GE HealthCare will correct all affected OEC 9800 and OEC 9900 Elite systems (28 units) by visually inspecting for the presence of lead shield tape on the collimator cover and replacing the collimator cover if it is found non-compliant, at no cost to the customer.
Quantity
not on file
Distribution
US Nationwide distribution.
Lot, serial or model codes
GTIN: 00840682114349
510(k) numbers
["K073543","K122234","K082781","K120613"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register