Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: MAGNETOM Aera magnetic resonance diagnostic devices Model # 10432914

Recall number
Z-2342-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
MAGNETOM Aera magnetic resonance diagnostic devices Model # 10432914
Status
Terminated
Initiated
2019-07-09
Posted
not on file
Terminated
2020-04-23
Reason
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
Root cause
Device Design
Action
1. review the System Owner Manual chapter 1.3.3 Quench emergency plan and the section regarding defective vent line. 2. follow the standard emergency procedures you have in place in the event of system failures. 3.A Siemens Service Technician will conduct an inspection to verify that your system has a vertical outlet. If a vertical outlet is present, no further action is required. In the event your system is equipped with a horizontal outlet, the Siemens Service Technician will ensure that the port for the drop-off plate is adequately blanked off or sealed. 4.service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. 5. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. By signing this document you are acknowledging that you have read and understand the content therein. If you have any question, please contact Meredith Adams, Telephone 610-448-6461 and E-Mail: Meredith.adams@siemens-healthineers.com
Quantity
143
Distribution
AL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY
Lot, serial or model codes
Serial # 41878 52409 142129 141526 42163 152267 141686 42436 42160 142132 41953 141226 141862 141521 42225 52171 41980 41051 152232 42247 141462 52299 52338 42416 141199 141557 142109 141054 152249 42020 41794 142147 152217 141951 42458 141638 42087 142010 42326 142009 142124 141676 42290 52229 42298 142130 141561 141973 42530 141141 41951 42236 152215 42267 42364 141305 42372 142122 42083 142142 42217 52195 152175 41945 141712 42256 42310 41780 42422 141963 42534 52350 52249 42412 42413 52397 41862 141786 42420 52221 52304 42306 141513 142024 141016 42495 142120 41906 52267 141589 42249 42158 41897 52321 42303 141056 42353 42330 41778 141275 141109 142107 42381 42339 41777 152234 52190 141201 42566 141677 41782 42254 42214 52192 141314 141585 42248 42295 42517 141072 42417 141439 141975 142104 41941 41813 141260 42533 141386 52157 141154 141135 142128 141121 142077 141242 142074 141253 141043 141221 42635 42507 141025
510(k) numbers
K111242
PMA numbers
not on file
Product code
LNH
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register