Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: N Cuvette Segments for BN ProSpec System-IVD a system for the quantitative determination of proteins in serum, plasma, urine and CSF. Siemens Material Number (SMN):10446539

Recall number
Z-2341-2019
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
N Cuvette Segments for BN ProSpec System-IVD a system for the quantitative determination of proteins in serum, plasma, urine and CSF. Siemens Material Number (SMN):10446539
Status
Terminated
Initiated
2019-06-27
Posted
not on file
Terminated
2020-05-27
Reason
Iincreased (falsely elevated) Albumin results compared to expected results
Root cause
Process change control
Action
Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-19-002.A.US to the US customers June 27, 2019 and May 2019 via FedEx overnight delivery and email to regional countries for implementation Outside the US (OUS). Customers were instructed to contact their Siemens Healthineers Customer Care Center for further follow-up in case they experienced sporadically increased Albumin results compared to expected results. Customers were also instructed to complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days.
Quantity
664 units
Distribution
Nationwide
Lot, serial or model codes
Lot # 5890128326 and Lot # 182670004 UDI: 00842768012334
510(k) numbers
["K001647"]
PMA numbers
not on file
Product code
JZW
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register