Intuitive Surgical, Inc.: da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Recall number
Z-2341-2018
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Mode IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Status
Terminated
Initiated
2018-03-07
Posted
not on file
Terminated
2020-07-29
Reason
Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
Root cause
Nonconforming Material/Component
Action
Intuitive Surgical, Inc. sent a Recall letters dated 3/7/2018 to affected customers via FedEx. The letter identified the affected product problem, risk to health, affected serial numbers, the action the firm will be taking, and the actions to be taken by the customer.
Quantity
108 ACJ's on 35 surgical systems
Distribution
US Nationwide AR, CA, FL, GA, IL, IA, MN, MS, NE, OK, TN, and TX. Foreign distribution was made to Australia, Austria, Finland, France, Germany, Italy, Netherlands, Spain, and United Kingdom.
Lot, serial or model codes
System S/N SL0008, ACJ S/N's P117154256, P117154263, P117154281, P117154286;  System S/N SL0012, ACJ S/N's P117154260, P117154297, P117154298,  P117154308;  System S/N SL0013, ACJ S/N's P117154251, P117154255, P117154257,  P117154305;  System S/N SL0015, ACJ S/N's P117154274, P117154276, P117154289,  P117154290;  System S/N SL0016, ACJ S/N's P117154259, P117154291, P117154306,  P117154307;  System S/N SL0017, ACJ S/N's P117154261, P117154269, P117154277,  P117154296;  System S/N SL0018, ACJ S/N's P117194009, P117194017, P117194037;  System S/N SL0019, ACJ S/N's P117154267, P117154283, P117154294,  P117154302;  System S/N SL0020, ACJ S/N's P117194008, P117194050, P117194060, P117194078;  System S/N SL0021, ACJ S/N's P117194063, P117194064, P117194081, P117194083;  System S/N SL0028, ACJ S/N's P117154279, P117154285, P117154292, P117154300;  System S/N SL0034, ACJ S/N's P117154264, P117154275, P117154304;  System S/N SL0035, ACJ S/N's P117194016, P117194019, P117194070, P117194093;  System S/N SL0036, ACJ S/N's P117194025, P117194073, P117241990;  System S/N SL0037, ACJ S/N's P117194006, P117194036, P117194056, P117194058  System S/N SL0038, ACJ S/N's P117241971, P117241993;  System S/N SL0039, ACJ S/N's P117194013, P117194024, P117194031, P117194032;  System S/N SL0040, ACJ S/N's P117194046, P117194061, P117194088, P117194094;  System S/N SL0041, ACJ S/N's P117194027, P117194040, P117194043, P117194049;  System S/N SL0042, ACJ S/N's P117154288, P117194015, P117194052, P117194090;  System S/N SL0043, ACJ S/N P117241988;  System S/N SL0044, ACJ S/N's P117194007, P117194085, P117241978;  System S/N SL0045, ACJ S/N's P117241965, P117241970, P117241982, P117241984;  System S/N SL0046, ACJ S/N's P117194014, P117241964, P117241976, P117241980;  System S/N SL0047, ACJ S/N's P117194020, P117194048, P117194067, P117194074  System S/N SL0048, ACJ S/N's P117194004, P117194072, P117200171;  System S/N SL0052, ACJ S/N's P117241977, P117241979, P117241986, P117241991;  System S/N SL0053, ACJ S/N P117241966;  System S/N SL0055, ACJ S/N P117283217;  System S/N SL0056, ACJ S/N P117283216;  System S/N SL0060, ACJ S/N P117241972;  System S/N SL0063, ACJ S/N P117241968;  System S/N SL0070, ACJ S/N's P117194028, P117194092;  System S/N SL0071, ACJ S/N's P117194018, P117283218; and  System S/N SL0073. ACJ S/N P117283214.
510(k) numbers
["K171294"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register