FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Artis Q systems; dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
- Recall number
- Z-2341-2015
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Artis Q systems; dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
- Status
- Terminated
- Initiated
- 2015-06-29
- Posted
- 2015-08-05
- Terminated
- 2016-12-07
- Reason
- The possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may no longer be possible. No injuries reported.
- Root cause
- Device Design
- Action
- A customer safety notice, dated June 29, 2015, was sent to direct accounts to alert them about the issue and that Siemens' service organization will be contacting them to perform the corrective action.
- Quantity
- not on file
- Distribution
- US Nationwide
- Lot, serial or model codes
- Model Number of device: 10848282 10848281 10848355 10848354 serial numbers: 121006 121013 109016 121011 123008 123003 123011 121014 109011 121004 121018 121036 121012 109015 109009 121019 121010 121015 121016 109007 111000 121020 121009 121056 121057 109021 121008 109020 109027 109010 123010 121041
- 510(k) numbers
- ["K123529"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28