Skip to the content

FDA Medical Device Recalls (openFDA)

W. L. Gore & Associates Inc.: Gore Molding & Occlusion Balloon Catheter, REF Catalogue Number MOB37, SN XXXXXXXX, 10 mm - 37 m 10 Fr, Guidewire Compatibility 0.035", STERILE EO, CE2797, UDI: (00)733132639489

Recall number
Z-2340-2021
Recalling firm
W. L. Gore & Associates Inc.
Product
Gore Molding & Occlusion Balloon Catheter, REF Catalogue Number MOB37, SN XXXXXXXX, 10 mm - 37 m 10 Fr, Guidewire Compatibility 0.035", STERILE EO, CE2797, UDI: (00)733132639489
Status
Open, Classified
Initiated
2021-07-12
Posted
not on file
Terminated
not on file
Reason
Complaints received concerning Balloon Catheter leakage from the guidewire lumen and the y-hub during preparation resulting in either insufficient inflation or an inability to inflate.
Root cause
Process change control
Action
On July 23, 2021 issue a "Urgent Medical Device Recall" notification to affected consignees via UPS. In addition to informing consignees about the recall the customer notification ask consignees to take the following actions: 1. Identify and return any unused devices within the scope of this voluntary recall. 2. Please complete and sign the enclosed CUSTOMER RESPONSE FORM and return to WLGore4731@stericycle.com within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. 4. If a listed device has been used, there is no patient-follow up needed and there are no further actions needed other than informing Gore the device was used. Please indicate the used devices on the CUSTOMER RESPONSE FORM and return to WLGore4731@stericycle.com within 2 weeks of receipt of this letter. No further action is needed.
Quantity
3,364 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AUS, CAN, CHE, DEU, DNK, ESP, FIN, FRA, GBR, GRC, HRV, HUN, IRL, ISR, ITA, JPN, MEX, NLD, NOR, POL, PRT, SMR, SWE.
Lot, serial or model codes
Serial Numbers between 22982051 - 23516741
510(k) numbers
["K172567"]
PMA numbers
not on file
Product code
MJN
Firm FEI
3007284313
Firm city
Phoenix
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register