FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
- Recall number
- Z-2339-2025
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
- Status
- Open, Classified
- Initiated
- 2025-07-25
- Posted
- 2025-08-19
- Terminated
- not on file
- Reason
- Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
- Root cause
- Component change control
- Action
- Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.
- Quantity
- 434 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Model Number: WA91307C. Lot Code: ESG-410 Legacy WA91307C. UDI-DI: 04042761086349. Serial Numbers: 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101023, 101024, 101025, 101029, 101030, 101031, 101032, 101033, 101034, 101035, 101036, 101037, 101038, 101039, 101040, 101041, 101042, 101043, 101044, 101045, 101046, 101047, 101048, 101049, 101050, 101051, 101052, 101053, 101054, 101055, 101056, 101057, 101058, 101059, 101060, 101061, 101062, 101063, 101064, 101066, 101067, 101068, 101069, 101070, 101071, 101072, 101073, 101074, 101238, 101239, 101240, 101241, 101242, 101243, 101244, 101245, 101246, 101247, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101260, 101261, 101262, 101263, 101264, 101265, 101266, 101267, 101268, 101269, 101270, 101271, 101272, 101273, 101274, 101275, 101276, 101277, 101278, 101279, 101280, 101281, 101282, 101283, 101284, 101285, 101286, 101287, 101288, 101289, 101290, 101291, 101292, 101293, 101294, 101295, 101296, 101297, 101298, 101299, 101300, 101301, 101302, 101333, 101334, 101335, 101336, 101337, 101338, 101339, 101340, 101341, 101342, 101370, 101371, 101372, 101373, 101374, 101375, 101376, 101377, 101378, 101379, 101395, 101396, 101397, 101398, 101399, 101400, 101401, 101402, 101403, 101432, 101433, 101434, 101435, 101436, 101437, 101438, 101439, 101440, 101441, 101613, 101633, 101634, 101635, 101636, 101637, 101638, 101684, 101685, 101686, 101687, 101688, 101689, 101691, 101690, 101692, 101693, 101694, 101695, 101743, 101744, 101745, 101746, 101747, 101748, 101749, 101750, 101751, 101752, 101753, 101754, 101755, 101756, 101757, 101758, 101759, 101760, 101761, 101762, 101763, 101764, 101765, 101766, 101767, 101768, 101769, 101770, 101771, 101772, 101773, 101774, 101775, 101776, 101777, 101778, 101779, 101780, 101781, 101782, 101783, 101784, 101785, 101786, 101787, 101788, 101789, 101958, 101959, 101960, 101961, 101962, 101963, 101964, 101965, 101966, 101967, 101968, 101969, 101970, 101971, 102017, 102018, 102019, 102020, 102021, 102024, 102025, 102026, 102027, 102028, 102055, 102056, 102057, 102058, 102059, 102060, 102061, 102062, 102063, 102064, 102065, 102066, 102067, 102068, 102069, 102284, 102285, 102286, 102287, 102288, 102289, 102290, 102291, 102292, 102293, 102294, 102295, 102296, 102297, 102298, 102299, 102300, 102301, 102302, 102303, 102304, 102305, 102306, 102307, 102308, 102309, 102310, 102311, 102312, 102313, 102314, 102315, 102316, 102317, 102318, 102319, 102320, 102321, 102323, 102325, 102326, 102327, 102328, 102329, 102330, 102331, 102332, 102333, 102334, 102335, 102362, 102363, 102364, 102366, 102367, 102368, 102369, 102370, 102371, 102372, 102373, 102391, 102392, 102393, 102394, 102395, 102396, 102397, 102399, 102400, 102401, 102402, 102403, 102404, 102405, 102406, 102407, 102408, 102409, 102410, 102411, 102412, 102413, 102414, 102415, 102416, 102417, 102418, 102419, 102420, 102421, 102422, 102423, 102424, 102425, 102426, 102428, 102429, 102430, 102431, 102432, 102433, 102434, 102435, 102436, 102437, 102438, 102439, 102440, 102441, 102442, 102443, 102444, 102445, 102497, 102498, 102499, 102500, 102501, 102502, 102503, 102504, 102505, 102506, 102507, 102508, 102509, 102510, 102511, 102512, 102513, 102515, 102516, 102517, 102518, 102519, 102520, 102521, 102522, 102523, 102802, 102807, 102808, 102809, 102810, 102814, 102818, 102819, 102820, 102821, 102822, 102823, 102828, 102829, 102830, 102831, 102832, 102834, 102836, 102837.
- 510(k) numbers
- ["K203277"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28