Skip to the content

FDA Medical Device Recalls (openFDA)

Hitachi Healthcare Americas Corporation: OASIS MRI System

Recall number
Z-2339-2021
Recalling firm
Hitachi Healthcare Americas Corporation
Product
OASIS MRI System
Status
Terminated
Initiated
2021-07-29
Posted
not on file
Terminated
2023-09-08
Reason
There was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is required by the Code of Federal Regulations section 21CFR801.437.
Root cause
Vendor change control
Action
On July 29, 2021, the firm distributed Urgent Medical Device Safety Alert letters to affected customers. Customers were informed that the anti-skid area on the bottom of the table pads contain dry natural rubber. If a patient or user inadvertently comes into contact with the anti-skid material, it may cause allergic symptoms. If this occurs, immediately seek medical care for the user or patient. The firm does not require any additional action by the user at this time. The firm will begin including revised labels and user manuals with newly manufactured systems which contain the required notification of the presence of dry natural rubber in the anti-skid area of the pads. The firm will not be taking any additional action for already distributed product. For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email aaron.pierce@fujifilm.com
Quantity
472 (US); 10 (OUS) Total
Distribution
Worldwide distribution - US Nationwide and the countries of Mexico and Brazil.
Lot, serial or model codes
M001-M329, M951
510(k) numbers
["K202030"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register