Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Toshiba Aquilion ONE CT SystemTSX-301 A/2D

Recall number
Z-2339-2010
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba Aquilion ONE CT SystemTSX-301 A/2D
Status
Terminated
Initiated
2009-02-03
Posted
2010-09-01
Terminated
2011-09-02
Reason
ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB.
Root cause
Software design
Action
Toshiba America Medical System Inc (TAMS) sent an Urgent: Medical Device Correction dated January 14, 2009 to their consignees. The letter described the issue and provided instructions to test the device to determine if it is an affected device. If the device is affected, contact the Toshiba service representative at 800 521-1968. Consignees were requested to complete and fax a response form to Toshiba.
Quantity
16
Distribution
Nationwide Distribution: including the states of AL, CA, FL, GA, KY, MA, MD, MT, NV, NY, OH, PA, and Puerto Rico.
Lot, serial or model codes
2DA0872038, 2DA0862034, 2DA07X2008, 2DA0842023, 2DA07X2005, 2DA0882053, 2DA0882051, 2DA0862036, 2DA0882049, 2DA0882052, 2DA0842026, 2DA07Y2010, 2DA0882048, 2DA08X2061, 2DA0792004, and 2DA0792002
510(k) numbers
["K072606"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register