Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: COULTER LH 500 Series System

Recall number
Z-2337-2010
Recalling firm
Beckman Coulter Inc.
Product
COULTER LH 500 Series System
Status
Terminated
Initiated
2010-02-12
Posted
2010-09-01
Terminated
2010-09-10
Reason
The recall was initiated after Beckman Coulter identified four (4) issues with the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System: Issue 1: There is a potential for misidentification to occur when the system is configured in languages other than English or Chinese. The characters #,@,[,\,],`,{,|,}, or ~ are substituted or omitted in other languages. The issu
Root cause
Software design
Action
The customer recall notification was initiated when Product Corrrective Actions letters with attached Customer Response form were sent on the week of February 15, 2010 to all customers who purchased the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System. The letters will provided the customers with an explanation of the problem identified and a work around. Customers with questions regarding the Product Corrective Action were instructed to call 800-526-7694 in the United States or Canada, or contact your local Beckman Coulter Representative.
Quantity
1373 units in US and 147 units in Canada
Distribution
Throughout the United States and Canada
Lot, serial or model codes
All serial Numbers.
510(k) numbers
["K032000"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register