Skip to the content

FDA Medical Device Recalls (openFDA)

Immunotech A.S.: Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.

Recall number
Z-2336-2025
Recalling firm
Immunotech A.S.
Product
Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.
Status
Open, Classified
Initiated
2025-06-12
Posted
2025-08-18
Terminated
not on file
Reason
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.
Root cause
Nonconforming Material/Component
Action
On 06/12/2025, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter determined that the Estrone RIA kit might be affected by an isolated coated tubes inhomogeneity issue causing falsely elevated results or falsely decreased results. Customers are instructed to: Beckman Coulter recommends sharing the content of the Recall Letter with their laboratory and/or medical director regarding the need to review previous patient test results. Note- All affected product have expired (expiry dates 12/31/2024 thru 04/22/2025) For questions, contact Customer Support Center: Website: http://www.beckmancoulter.com Email: imunochem@beckman.com
Quantity
279 units
Distribution
Worldwide - U.S. Nationwide distribution in the states of NC and the countries of Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, and Netherlands.
Lot, serial or model codes
Catalog Number DSL8700 UDI Code: 15099590211615 Serial/Lot Numbers: 241014D 241111C 241209C 241209D 250106C 250106D 250203C
510(k) numbers
["K935013"]
PMA numbers
not on file
Product code
CGF
Firm FEI
3003385757
Firm city
Prague 10
Firm state
not on file
Firm country
Czech Republic

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register