Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Recall number
Z-2335-2012
Recalling firm
Steris Corporation
Product
The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
Status
Terminated
Initiated
2012-08-01
Posted
2012-09-07
Terminated
2014-06-06
Reason
The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.
Root cause
Nonconforming Material/Component
Action
STERIS sent a Urgent Voluntary Field Correction Notice letter dated August 1, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. STERIS has received no reports of injuries associated with the disconnection of a SYSTEM IE water hose. STERIS Corporation will instill new hoses and connections on your SYSTEM IE Liquid Chemical Sterilant Processing System. For further information or if you have any questions regarding our visit to your facility, please contact STERIS Field Service Dispatch at 1-800-333-8828.
Quantity
6,051 units
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Bahrain and Hong Kong.
Lot, serial or model codes
Model #: P6500, S/N 400000 through 406294
510(k) numbers
["K090036","K101409","K102462"]
PMA numbers
not on file
Product code
MED
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register