FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Triathlon Total Knee System Primary Tibial Baseplate; Sterile, Made in Ireland. Howmedica Osteonics 325 Corporate Drive Mahwah, NJ 07430 The Triathlon system is comprised of a femoral component, tibial tray, and tibial insert that are intended to be used in total knee arthroplasty (if replacement of the articular surf ace of the patella is required, the Duracon patellar components are compatible with the Triathlon components). The Triathlon PS Total Knee System is intended to be used in situation where the posterior cruciate ligament is absent, or cannot be preserved. The Triathlon CR Total Knee System consists of a femoral component, tibial insert, and all-polyethylene patellar components that are intended to be used with the previously cleared Triathlon Primary Cemented Tibial Tray in primary or revision total knee arthroplasty. The Triathlon All Polyethylene patellare components are intended to be used with the femoral components of the previously released Duracon Total Knee System, as well as the previously release Triathlon PS Femoral component in situation where replacement of the articular surface of the patella is required. The Triathlon CR Total Knee system is intended to accommodate the posterior cruciate ligament (PCL) if it is present.
- Recall number
- Z-2334-2010
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Triathlon Total Knee System Primary Tibial Baseplate; Sterile, Made in Ireland. Howmedica Osteonics 325 Corporate Drive Mahwah, NJ 07430 The Triathlon system is comprised of a femoral component, tibial tray, and tibial insert that are intended to be used in total knee arthroplasty (if replacement of the articular surf ace of the patella is required, the Duracon patellar components are compatible with the Triathlon components). The Triathlon PS Total Knee System is intended to be used in situation where the posterior cruciate ligament is absent, or cannot be preserved. The Triathlon CR Total Knee System consists of a femoral component, tibial insert, and all-polyethylene patellar components that are intended to be used with the previously cleared Triathlon Primary Cemented Tibial Tray in primary or revision total knee arthroplasty. The Triathlon All Polyethylene patellare components are intended to be used with the femoral components of the previously released Duracon Total Knee System, as well as the previously release Triathlon PS Femoral component in situation where replacement of the articular surface of the patella is required. The Triathlon CR Total Knee system is intended to accommodate the posterior cruciate ligament (PCL) if it is present.
- Status
- Terminated
- Initiated
- 2009-09-01
- Posted
- 2010-08-31
- Terminated
- 2012-08-27
- Reason
- A lot for lot mix up occurred between the Triathlon Total Knee System Primary Tibial Baseplate # 4 Cemented, Catalog # 5520-B-400, Lot SH8NT and the #2 Cemented, Catalog # 5520-B-200, Lot SH7RP.
- Root cause
- Mixed-up of materials/components
- Action
- The firm, Stryker Orthopaedics, sent an "URGENT PRODUCT RECALL" letter dated September 1, 2009, via FedEx to all customers. The letter described the product, problem and action to be taken by customers. The customers were instructed to examine their inventory and hospital locations to identify the affected product; retrieve and return all affected products via Inter-Org Transfer and ship to Stryker Orthopaedics, 325 Corporate Drive, Mahwah, New Jersey, 07430; Attn: Regulatory Reporting, and attach the fluorescent orange PRODUCT REMEDIATION sticker and mark the outer box with the words "PRODUCT RECALL; complete and fax the attached Product Accountability Form and spreadsheet to (201) 831-6069 within 5 days of receipt of the notice. If you have any questions, please contact the Manager, Divisional Regulatory Reporting at (201) 831-5970.
- Quantity
- Lot Code: SH7RP - 6 in the US. Lot Code SH8NT was not received in the US.
- Distribution
- Nationwide distribution: AL, AZ, CA, MT, NJ and TX.
- Lot, serial or model codes
- # 4 - Catalog Number: 5520-B-400, Lot code: SH8NT; # 2 - Catalog Number: 5520-B-200, Lot Code SH7RP.
- 510(k) numbers
- ["K031729","K040267"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28