Skip to the content

FDA Medical Device Recalls (openFDA)

Aesculap Implant Systems LLC: Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities: 1 - degenerative disc disease, 2 - spondylolisthesis, 3 - trauma, 4 - spinal stenosis, 5 - deformities or curvatures, 6 - tumor, 7 - pseudoarthrosis, and 8 - failed previous fusion.

Recall number
Z-2333-2012
Recalling firm
Aesculap Implant Systems LLC
Product
Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities: 1 - degenerative disc disease, 2 - spondylolisthesis, 3 - trauma, 4 - spinal stenosis, 5 - deformities or curvatures, 6 - tumor, 7 - pseudoarthrosis, and 8 - failed previous fusion.
Status
Terminated
Initiated
2012-07-25
Posted
2012-09-07
Terminated
2014-03-18
Reason
There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert preventing the screw head from locking down to the bone screw.
Root cause
Nonconforming Material/Component
Action
Aesculap sent an "IMPORTANT PRODUCT REMOVAL NOTIFICATION" letter dated July 25, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Distribution Inventory Sheet was attached for customers to complete and return with the affected product. Contact the firm at (610) 984-9275 for questions regarding this recall.
Quantity
5
Distribution
Nationwide Distribution-including the states of IL, MN, NV, OK, and TX.
Lot, serial or model codes
All distributed lots and serial numbers
510(k) numbers
["K100623"]
PMA numbers
not on file
Product code
NKB
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register