FDA Medical Device Recalls (openFDA)
Oscor, Inc.: Permanent Pacing Lead (Catalog #: 03332, 03334, 03337, 03107, 03116, 03108, 03109, 05756, 53420, 05765, 03634, 03633, and 05747); Model #: PY244RU, PY252RU, PY258RU, REFINO58RU, REFINO 52RJU, REFINO 44RJU, REFINO48RJU, BIS/IS-15, BIS/BIS-17, VKU-10V, MP52PV, MP40PV, C/IS-10; Manufactured by Oscor. Pacing and sensing of the ventricle and/or atrium of the heart in conjunction with a compatible, implantable pulse generator/pacemaker.
- Recall number
- Z-2333-2008
- Recalling firm
- Oscor, Inc.
- Product
- Permanent Pacing Lead (Catalog #: 03332, 03334, 03337, 03107, 03116, 03108, 03109, 05756, 53420, 05765, 03634, 03633, and 05747); Model #: PY244RU, PY252RU, PY258RU, REFINO58RU, REFINO 52RJU, REFINO 44RJU, REFINO48RJU, BIS/IS-15, BIS/BIS-17, VKU-10V, MP52PV, MP40PV, C/IS-10; Manufactured by Oscor. Pacing and sensing of the ventricle and/or atrium of the heart in conjunction with a compatible, implantable pulse generator/pacemaker.
- Status
- Terminated
- Initiated
- 2008-01-29
- Posted
- 2008-09-24
- Terminated
- 2008-11-14
- Reason
- The O-rings are over tolerance making it hard to connect the leads to the pacemaker.
- Root cause
- Nonconforming Material/Component
- Action
- On January 24, 08, Oscor sent out a letter to doctors/ customers with the subject line of "Re: Market Withdrawal of Permanent Pacing Leads due to the lead connector insertion difficulty," stating they experienced some difficulties with the insertion of pacing lead connectors into the pacemaker headers. They asked that customers review the inventory list and remove all products, whose serial numbers match the serial numbers on the list and have not been implanted. Also, customers should call Oscor customer service for a Return Goods Product number and return the product back to the firm for replacement.
- Quantity
- not on file
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- 02-01502, 02-01812, 02-01436, 02-01440, 02-02016, and 02-01945
- 510(k) numbers
- ["K010787","K040569","K041809","K872971"]
- PMA numbers
- not on file
- Product code
- DTB
- Firm FEI
- 1035166
- Firm city
- Palm Harbor
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28