Skip to the content

FDA Medical Device Recalls (openFDA)

Sunquest Information Systems, Inc.: Sunquest Laboratory Sunquest Laboratory is intended for use by professionals working in a clinical laboratory. Sunquest Laboratory is intended for use only by professionals who have received extensive training in the use of the software. Sunquest Laboratory is controlled via userdefined security levels, which allow the laboratories to manage data security and integrity. The system is marketed as an automated solution for managing laboratory processes, including: " Patient registration " Order entry and order modification " Specimen collection, verification, suitability, and distribution " Quality assurance checking " Workload recording " Billing charge capture " Standards and controls recording " Test order result inquiry and reporting Organism susceptibility and epidemiology records " Microbiology culture direct examination and observation recording " Outreach clients, specimens, and results " Client-specific reporting capabilities " Inbound client service request tracking " Historical data Clients that use the Sunquest Blood Bank system should refer to the intended use statement found in the Blood Bank System Overview Users Guide.

Recall number
Z-2332-2012
Recalling firm
Sunquest Information Systems, Inc.
Product
Sunquest Laboratory Sunquest Laboratory is intended for use by professionals working in a clinical laboratory. Sunquest Laboratory is intended for use only by professionals who have received extensive training in the use of the software. Sunquest Laboratory is controlled via userdefined security levels, which allow the laboratories to manage data security and integrity. The system is marketed as an automated solution for managing laboratory processes, including: " Patient registration " Order entry and order modification " Specimen collection, verification, suitability, and distribution " Quality assurance checking " Workload recording " Billing charge capture " Standards and controls recording " Test order result inquiry and reporting Organism susceptibility and epidemiology records " Microbiology culture direct examination and observation recording " Outreach clients, specimens, and results " Client-specific reporting capabilities " Inbound client service request tracking " Historical data Clients that use the Sunquest Blood Bank system should refer to the intended use statement found in the Blood Bank System Overview Users Guide.
Status
Terminated
Initiated
2012-03-15
Posted
2012-09-07
Terminated
2012-10-02
Reason
Sunquest is recalling Sunquest Laboratory versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1 because when the Windows-based merge application is used and the "Event merge across patients" option is used, a receiving system (for example a HIS or CoPath system) may interpret the event merge as a full patient merge. This occurs because a trigger from the Laboratory system for the event
Root cause
Software design
Action
The firm, Sunquest, sent an "Urgent- Product Safety Notice: PSN-12-02" dated March 15, 2012 to all their customers who purchased the Sunquest Laboratory versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1. The notice described the product, the problems and the actions to be taken. Customers were instructed to request the software correction by creating a Case via the Client Support webpage at: www.sunquestinfo.com/supportweb. Customers with questions were instructed to contact Client support at (877) 239-6337 for US and Canadian Sites and at +44 (0) 161 335 0562 for international sites.
Quantity
126 sites
Distribution
Worldwide distribution: USA (nationwide) and countries of: England, United Kingdom and United Arab Emirates.
Lot, serial or model codes
Version versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000306472
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register