Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, LP: AltiVate Anatomic Shoulder System for the following Part Numbers: 1. Part # 520-08-00, Rev. C., 2. Part # 520-10-000, 3. Part # 520-12-000, 4. Part # 520-14-000, 5. Part # 520-16-000

Recall number
Z-2331-2021
Recalling firm
Encore Medical, LP
Product
AltiVate Anatomic Shoulder System for the following Part Numbers: 1. Part # 520-08-00, Rev. C., 2. Part # 520-10-000, 3. Part # 520-12-000, 4. Part # 520-14-000, 5. Part # 520-16-000
Status
Open, Classified
Initiated
2021-06-28
Posted
not on file
Terminated
not on file
Reason
Complaints have been received concerning the humeral stem failing to mate with the broach handles for a shoulder device. This could result in delay of procedure or revision surgery.
Root cause
Process change control
Action
On June 28, 2021, Encore Medical (DJO) issued a "Urgent Medical Device Recall" Notification to affected consignees, via E-Mail. In addition to informing consignees about the recall devices, the firm asked consignees to take the following actions by July 19, 2021: 1. Review your stock for the affected products and lot numbers and quarantine the items until returned. 2. Communicate the affected parts, lots, and quantities that you have in stock to your sales Agent and arrange to return all affected, unused product to your sales Agent. 3. Report any adverse events related to this issue to DJO Customer Service (1-800-456-8696). Please also note that Adverse reactions or quality problems experienced with the use of this product MAY be reported to the FDAs MedWatch Adverse Event reporting program either online, by regular mail or by fax. 4. Complete the Medical Device Recall acknowledge and receipt form (including all information in the yellow-highlighted cells) with all available information and return by e-mail to julie.baker@djoglobal.com and david.evans1@djoglobal.com by July 31, 2021.
Quantity
not on file
Distribution
US: AL,AR,AZ,CA,CO,FL,GA,HI,IL,IN,KY,LA,MD,ME,MI,MN,MO,MS,MT,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,TN,TX,UT,VA,WA,WV OUS: N/A
Lot, serial or model codes
Part number/ Lot number:   1. 520-08-000 / 85OU1108; 85OU1110; 85OU1111,  2. 520-10-000 / 851U1256; 851U1257; 851U1258; 851U1259,  3. 520-12-000 / 852U1402; 852U1393; 852U1404; 852U1395; 852U1398A, 852U1405; 852U1397; 852U1391; 852U1400; 852U1398; 852U1407; 852U1396; 852U1406,  4. 520-14-000 / 853U1234; 853U1240; 853U1241A; 853U1238; 853U1239; 853U1237; 853U1241,  5. 520-16-000 / 854U1076; 854U1077
510(k) numbers
["K162024"]
PMA numbers
not on file
Product code
KWS
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register