Skip to the content

FDA Medical Device Recalls (openFDA)

Propper Mfg Co Inc: Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manufacturing Co., Inc., Long Island City, N.Y. 11101

Recall number
Z-2330-2008
Recalling firm
Propper Mfg Co Inc
Product
Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manufacturing Co., Inc., Long Island City, N.Y. 11101
Status
Terminated
Initiated
2008-02-07
Posted
2008-09-20
Terminated
2012-08-03
Reason
The firm discovered a production specification discrepancy that may cause the indicators to show an inaccurate result.
Root cause
Equipment maintenance
Action
On 2/07/08, Propper Mfg. Co. sent URGENT: DEVICE RECALL letters (dated 2/07/08) to the direct consignees, informing them of the recall of the affected lots of Gas-Chex EO Indicator Strips. The letters also instructed them to: (1) check their inventory to determine if they have the recalled lots in stock; (2) discontinue use of the affected lots; (3) reprocess any unused packs that have been processed with the affected lots of Gas-Chex EO Indicator Strips with non-recalled lots or an alternative product from another supplier; (4) return the affected lots to Propper Manufacturing; (5) complete the enclosed form and return via fax to Propper Mfg. Propper Mfg. will send a postage-paid label and instruction for return of the product, and will ship replacement product as soon as possible. Propper also recommended that consignees review the risk associated with items that have been monitored with the subject chemical indicators although a sterilization cycle failure is unlikely due to equipment controls and other monitoring methods. For questions, please contact Propper at 1-800-832-4300 during regular business hours (9 am to 4:30 pm).
Quantity
1757 cases (7030 boxes)
Distribution
Nationwide and Puerto Rico
Lot, serial or model codes
Reorder No. 269001; Lots 6081, 6043, 6021, 7042, and 7013
510(k) numbers
not on file
PMA numbers
not on file
Product code
JOJ
Firm FEI
2410251
Firm city
City
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register