Skip to the content

FDA Medical Device Recalls (openFDA)

AccessClosure, Inc., A Cardinal Health Company: 6F/7F MynxGrip Vascular Closure Device; Product Model MX6721, Product Usage: The 6F/7F MynxGrip Vascular Closure Device is indicated for use to seal femoral arterial and femoral venous access sites while reducing times to hemostasis and ambulation in patients who have undergone diagnostic or interventional endovascular procedures utilizing a 6F or 7F procedural sheath. The 6F/7F MynxGrip Vascular Closure Device is designed to achieve femoral artery and femoral vein hemostasis via delivery of an extravascular, water-soluble synthetic sealant using a balloon catheter in conjunction with a standard procedural sheath. The sealant is comprised of a polyethylene glycol (PEG) material which expands upon contact with subcutaneous fluids to seal the arteriotomy or venotomy. The sealant is resorbed by the body within 30-days.

Recall number
Z-2329-2016
Recalling firm
AccessClosure, Inc., A Cardinal Health Company
Product
6F/7F MynxGrip Vascular Closure Device; Product Model MX6721, Product Usage: The 6F/7F MynxGrip Vascular Closure Device is indicated for use to seal femoral arterial and femoral venous access sites while reducing times to hemostasis and ambulation in patients who have undergone diagnostic or interventional endovascular procedures utilizing a 6F or 7F procedural sheath. The 6F/7F MynxGrip Vascular Closure Device is designed to achieve femoral artery and femoral vein hemostasis via delivery of an extravascular, water-soluble synthetic sealant using a balloon catheter in conjunction with a standard procedural sheath. The sealant is comprised of a polyethylene glycol (PEG) material which expands upon contact with subcutaneous fluids to seal the arteriotomy or venotomy. The sealant is resorbed by the body within 30-days.
Status
Terminated
Initiated
2016-07-13
Posted
2016-08-03
Terminated
2016-08-15
Reason
A total of 70 6F/7F Mynx Grip devices were shipped from a distributor using improper transport/packaging configurations.
Root cause
Process control
Action
Cardinal Health sent an Urgent Medical Device letters dated July 15, 2016 to affected customers. The letters provided the reason for recall and the exact lot number and quantity shipped to each location. Customers were instructed to quarantined affected products and complete and return the Field Action response form indicating the number of devices that will be returned. Instructions for returning the devices are included with an RMA number and Fed Ex airbill. Questions should be directed to Customer Advocacy at 408-610-6570 between 9 am and 5 pm PST.
Quantity
55 units
Distribution
US in the states of CA, and MS.
Lot, serial or model codes
Lot No: F1609702, Expiry 4/20/2018; Lot No. F1534302, Expiry 12/31/2016
510(k) numbers
not on file
PMA numbers
["P040044"]
Product code
MGB
Firm FEI
3010142363
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register