SCC Soft Computer: SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing specimen reports, billing and management reports for surgical, GYN, NON-GYN and Autopsy cases.

Recall number
Z-2329-2008
Recalling firm
SCC Soft Computer
Product
SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing specimen reports, billing and management reports for surgical, GYN, NON-GYN and Autopsy cases.
Status
Terminated
Initiated
2006-02-23
Posted
2008-09-22
Terminated
2008-11-20
Reason
Loss of text misrepresented individual tissue diagnosis on the final diagnosis print out from the HIS system software. The specimen heading "J" was removed and the diagnosis appeared under the specimen heading "I" as a result the specimen with the heading "I" had two diagnosis.
Root cause
Other
Action
Software corrections were sent as of 10/30/2003 for clients to upgrade to, the correction software patch is in versions 3.17.5.16, 3.17.6.10, 3.17.7.3 and 4.1.1.0. The firm did not inform the non-affected consignees of the failure because the firm concluded that the associated risk was low.
Quantity
335
Distribution
Worldwide Distribution including USA and the country of Canada.
Lot, serial or model codes
Versions 3.17 and 4.1
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000307651
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register