Skip to the content

FDA Medical Device Recalls (openFDA)

TITAN SPINE, LLC: Titan Spine Endoskeleton Locking Bone Screw Assembly, which is part of the Endoskeleton TCS Interbody Fusion Device, spinal implant. Part Numbers / Product Description: 5302-3514 / 03.5 x 14mm Locking Screws; 5302-3516 / 03.5 x 16mm Locking Screws; 502-3518 / 03.5 x 18mm Locking Screws; 5302-3814 / 03.8 x 14mm Locking Screws; 5302-3816 / 03.8 x 16 mm Locking Screws; 5302-3818 / 03.8 x 18mm Locking Screws.

Recall number
Z-2327-2015
Recalling firm
TITAN SPINE, LLC
Product
Titan Spine Endoskeleton Locking Bone Screw Assembly, which is part of the Endoskeleton TCS Interbody Fusion Device, spinal implant. Part Numbers / Product Description: 5302-3514 / 03.5 x 14mm Locking Screws; 5302-3516 / 03.5 x 16mm Locking Screws; 502-3518 / 03.5 x 18mm Locking Screws; 5302-3814 / 03.8 x 14mm Locking Screws; 5302-3816 / 03.8 x 16 mm Locking Screws; 5302-3818 / 03.8 x 18mm Locking Screws.
Status
Terminated
Initiated
2015-07-20
Posted
2015-08-03
Terminated
2015-10-20
Reason
Titan Spine is recalling specific Endoskeleton TCS Integrated Locking Bone Screw Assembly due to the locking bone screw collars not 'locking'.
Root cause
Nonconforming Material/Component
Action
Physicians and Distributors were sent on 7/20/2015 separate Titan "Urgent Medical Device Recall" letters dated July 20, 2015. The letters described the problem and the product involved in the recall. Advised consignees to complete and return the enclosed response form. For questions they contact Jane Rodd Senior AQ/RA Manager at 262-242-7801. Distributors were advised to examine their inventory and quarantine the product. If they have further distributed the product they should identify their customers and advise Titan Spine of their contact information.
Quantity
137 units
Distribution
Distributed in the states of CA, TX, OK, MA, NY, IN, OR, and GA.
Lot, serial or model codes
Lot numbers: V03, V05, V06
510(k) numbers
["K142940"]
PMA numbers
not on file
Product code
OVE
Firm FEI
3006340236
Firm city
Mequon
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register