Skip to the content

FDA Medical Device Recalls (openFDA)

INSPIREMD Inc: Brand Name: CGuard Prime Carotid Stent System, 135cm, 8mmx40mm Model/Catalog Number: CND0840

Recall number
Z-2326-2026
Recalling firm
INSPIREMD Inc
Product
Brand Name: CGuard Prime Carotid Stent System, 135cm, 8mmx40mm Model/Catalog Number: CND0840
Status
Open, Classified
Initiated
2026-05-01
Posted
2026-06-05
Terminated
not on file
Reason
Stent delivery system may encounter complications during deployment which may result in high resistance or inability to deploy the stent.
Root cause
Component design/selection
Action
On May 1, 2026, InspireMD, Inc. issued a Urgent Medical Device Recall notification to affected consignees. InspireMD asked consignees to take the following actions: 1. Cease further use of all affected CGuard Prime Carotid Stent Systems identified by the model numbers listed in the letter. 2. Quarantine and segregate all affected units currently in your inventory. Place affected product in a clearly labeled quarantine area to prevent inadvertent use. 3. This notice needs to be passed on to all those who need to be aware within your organization or facilities where the affected devices have been transferred. 4. Please retain a copy of this letter, and keep a copy of the Acknowledgement Form. 5. Complete the attached Acknowledgement Form and return it by email to USSupport@inspiremd.com within 2 business days. 6. InspireMD will contact each affected customer to facilitate the return and disposition process.
Quantity
424 units
Distribution
US Nationwide distribution in the states of AR, AZ, CA, DC, FL, GA, HI, IL, IN, KS, MA, MD, MI, MO, NC, NJ, NV, NY, OH, OR, RI, SD, TN, TX, VA, WA, WI, WV.
Lot, serial or model codes
UDI: 07290120281707; all lots/serial numbers manufactured between 02/03/2025 and 02/23/2026
510(k) numbers
not on file
PMA numbers
["P240029"]
Product code
NIM
Firm FEI
3032814119
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register