Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296

Recall number
Z-2326-2025
Recalling firm
Boston Scientific Corporation
Product
ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
Status
Open, Classified
Initiated
2025-07-24
Posted
2025-08-20
Terminated
not on file
Reason
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Root cause
Component design/selection
Action
On July 24, 2025, firm notified physicians via Important Medical Device Information letters. The letter provides information on the management of gradually rising daily subthreshold, low-voltage shock impedance (LVSI) pattern associated with calcification of expanded polytetrafluoroethylene (ePTFE) coated single coil (SC) and dual coil (DC) RELIANCE defibrillation leads manufactured by Boston Scientific Corporation between 2002 to 2021. Letter includes recommendations for routine follow-ups of affected leads, and considerations to take if lead replacement is planned. The firm is not recommending any changes to the scheduled follow-up interval for patients with ePTFE lead models.
Quantity
247,702 US; 344,670 OUS
Distribution
Worldwide Distribution. US nationwide. International distribution worldwide.
Lot, serial or model codes
Model 0282 UDI-DI/GTIN 00802526431371; Model 0283 UDI-DI/GTIN 00802526431388 00802526504181; Model 0285 UDI-DI/GTIN 00802526431401; Model 0286 UDI-DI/GTIN 00802526431418; Model 0292 UDI-DI/GTIN 00802526431432 00802526474668 00802526480096 00802526485763 00802526504396 00802526531248 00802526541742 00802526552441 00802526611186; Model 0293 UDI-DI/GTIN 00802526431449 00802526531255 00802526541827 00802526552526 00802526554490 00802526611179; Model 0295 UDI-DI/GTIN 00802526431463 00802526474699 00802526531262 00802526541902 00802526554506 00802526611131; Model 0296 UDI-DI/GTIN 00802526431470 00802526474705 00802526480164 00802526480188 00802526531279 00802526541988 00802526552038
510(k) numbers
not on file
PMA numbers
["P910073"]
Product code
NVY
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register