FDA Medical Device Recalls (openFDA)
Canon Medical System, USA, INC.: Infinix-8000V with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
- Recall number
- Z-2325-2020
- Recalling firm
- Canon Medical System, USA, INC.
- Product
- Infinix-8000V with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
- Status
- Open, Classified
- Initiated
- 2020-03-03
- Posted
- not on file
- Terminated
- not on file
- Reason
- Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
- Root cause
- Process control
- Action
- On 03/03/20, Urgent Medical Device Correction notices were mailed to Radiology Managers/Directors. To correct this issue, the bolts holding the longitudinal movement of the table will need to be inspected to determine if the bolts are tightened to specification. Recalling firm service representatives will contact customers to set up appointments to inspect bolts. Radiology Managers/Directors were advised of the following: a) Should this problem occur before your appointment with the recalling firm's service representative, please stop using the system and contact the service representative. b) If abnormalities are found operating the table during a procedure, please stop use, and contact the service representative. c) Share this information with all users and reviewing radiologists as well as clinical engineering, and the biomedical group at your facility. d) Complete and return the Customer Reply Form. Additional questions should be directed to (800) 421-1968 or your local service representative (800) 521-1968
- Quantity
- 46
- Distribution
- US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.
- Lot, serial or model codes
- Serial Numbers: A5B1642006, A1B17X2004, A5A1612005, A5B1772009, B5B1962002, A3A1522004, A7B16Y2001, A3A1712011, A5B1792010, A5A1532002, A1A1612003, A3A15Z2007, A5A1552003, A5B16Y2007, A3B1862016, A3B1812013, A3A1662009, A9A16X2004, BTA1922001, A3B1842015, AOB1882001, BMB1942001, AOB1892002, A5B1472001, A5B17Z2011, A3A14X2002, A5A1592004, A3A17Z2012, A5C1832012, A5B16Z2008, A1A1582002, A9A1582002, A1A14X2001, A3A1512003, A7C1812002, ABC18X2002, A3A1592005, UTC14Y2008, A3A15X2006, B5B1942001, A3A1682010, A1C18Z2005, A9A1662003, A3A1492001, A3B1822014, A3A1622008
- 510(k) numbers
- ["K170909"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28