Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

Recall number
Z-2323-2024
Recalling firm
Baxter Healthcare Corporation
Product
Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.
Status
Open, Classified
Initiated
2024-06-03
Posted
2024-07-10
Terminated
not on file
Reason
During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
Root cause
Employee error
Action
On 06/03/2024, Baxter Healthcare's Customer Service team contacted the consignee, via telephone call, to discuss the recall action and request the affected unit be returned to Baxter.
Quantity
1 unit
Distribution
Distribution to Illinois.
Lot, serial or model codes
UDI/DI: 00732094177640. Product Code: CP150A-1ENB. Serial Number: 100013892024
510(k) numbers
K141582
PMA numbers
not on file
Product code
DPS
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source