Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Medtronic Patient programmer, model 37742. The Patient programmer is a hand held, battery operated, microcontroller based device for use by patients to control and monitor external and implantable devices. The minimum amount of control involves the ability to tun therapy ON or OFF; maximal control allows the patient to select from different therapies and adjust the parameters with in the set therapy limits.

Recall number
Z-2323-2008
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Patient programmer, model 37742. The Patient programmer is a hand held, battery operated, microcontroller based device for use by patients to control and monitor external and implantable devices. The minimum amount of control involves the ability to tun therapy ON or OFF; maximal control allows the patient to select from different therapies and adjust the parameters with in the set therapy limits.
Status
Terminated
Initiated
2008-05-12
Posted
2008-09-20
Terminated
2008-11-21
Reason
Medtronic is retrieving four (4) patient programmers that were not properly loaded with application software. The application software is needed for a patient programmer to synchronize and bond with a neurostimulation device. Without this functionality a patient programmer is not useable and can not communicate with a neurostimulation device.
Root cause
Process control
Action
A letter was sent to each account telling them of the issue and to let them know that Medtronic is retrieving the affected devices. Medtronic Reps were to visit affected accounts to notify them of the recall, locate the affected device and return it. Rep is to complete the reply card and fax to Medtronic Neuromodulation. If you have questions, contact your Medtronic representative or Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Quantity
2
Distribution
GA, MO, NC, TX no OUS distribution
Lot, serial or model codes
serial numbers: NJD058707N, NJD062024N, NJD063259N, and NJD063324N
510(k) numbers
not on file
PMA numbers
["P840001S086"]
Product code
LGW
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register