Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc.: Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;

Recall number
Z-2321-2026
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
Status
Open, Classified
Initiated
2026-05-12
Posted
2026-06-05
Terminated
not on file
Reason
Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
Root cause
Process design
Action
On 5/12/2026 an IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers to inform them that manipulation of the luer-lock location during treatment preparation may result in fluid spillage onto the user and floor. This occurs prior to treatment and renders the product unusable. Direct contact with the fluid does not present any risk to the user as frangible manipulation only occurs after the peal seam has been broken and the two fluid compartments have been mixed. The potential risk associated with the affected Bicarby Dialysate product is a slip and fall hazard due to the presence of fluid on the floor.
Quantity
1,756,108 units
Distribution
US Nationwide distribution in the states of NY, TN and TX.
Lot, serial or model codes
Model Number: RFP-400-G; UDI-DI: 00840861102839; 2. Model Number: RFP-401-G; UDI-DI: 00840861102853; 3. Model Number: RFP-402-G; UDI-DI: 00840861102822; 4. Model Number: RFP-404-G; UDI-DI: 00840861102914; 5. Model Number: RFP-407-G; UDI-DI: 00840861102846: 6. Model Number: RFP-456-G; UDI-DI: 00840861102945; All lots produced from January 13, 2025, through lot 26AG02019 on January 13, 2026
510(k) numbers
["K233950","K243786"]
PMA numbers
not on file
Product code
KPO
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register