Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries Inc: Medline RF Detect Sterile X-Ray Detectable USP Type VII Gauze; 8" x 4", 12-ply, 100% cotton sponges, 10 sponges per pack, 80 packs per case; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA, Made in China; reorder: NON21432RF

Recall number
Z-2321-2008
Recalling firm
Medline Industries Inc
Product
Medline RF Detect Sterile X-Ray Detectable USP Type VII Gauze; 8" x 4", 12-ply, 100% cotton sponges, 10 sponges per pack, 80 packs per case; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA, Made in China; reorder: NON21432RF
Status
Terminated
Initiated
2008-06-13
Posted
2008-09-20
Terminated
2009-01-09
Reason
There is the potential for the RF ID tag to separate from the retaining pouch.
Root cause
Device Design
Action
Initially a recall letter dated 6/13/08 was sent to the direct accounts identifying 5 lots of affected lots with two reorder numbers involved, explaining that there was a lack of integrity of the retaining pouch material for the RFID gauze, and that there was a potential for the RFID tag to separate from the retaining pouch during use. It was subsequently discovered that there was only one reorder number affected by the problem and that an additional lot had been gamma sterilized. Therefore a second letter dated 6/19/08, listing the correct reorder number and the six affected lots, was sent via FedEx overnight on the same date. The accounts were requested to remove RFID gauzes with the listed lot numbers from all stock rooms, materials storage, and OR's, and return it to Medline for replacement. The accounts were requested to complete the enclosed response sheet and fax it back to Medline at 847-643-4482, indicating the amount and lot numbers of the gauzes being returned. Any questions were directed to 866-359-1704, between 8:00 AM and 5:00 PM CST. Distributors were requested to notify their accounts of the recall. If you have questions about this recall, contact Kathy Dunne at 847-970-4678 or at kdunne@medline.com.
Quantity
295 cases
Distribution
California, Indiana, Missouri, Ohio, Oklahoma, Pennsylvania, Texas and Washington
Lot, serial or model codes
reorder: NON21432RF, lots CM130002052, CM130002062, CM130002075, CM130002068, CM130002079, and CM130002110
510(k) numbers
["K061316"]
PMA numbers
not on file
Product code
GDY
Firm FEI
1417592
Firm city
Mundelein
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register