FDA Medical Device Recalls (openFDA)
Customed, Inc: CESAREAN PACK - - (7) TOWELS ABSORBENT 15" X 20" (2) SCALPEL WEIGHTED SAFETY #1 0 (1) BLANKET BABY 30" X 30" POLY ABS IMP (1) BOWL UTILITY QUART 32oz. (1) TUBE FEEDING 8Fr. 15" LONG (2) COUNTER NEEDLE/BLADE 10C MAG/CLEAR (1) BEANNIES BABY (2) PAD OBSTRETICAL XLGE ST. (1) SYRINGE 20cc W/0 NEEDLE LIL (1) SYRINGE EAR/ULCER 2oz. (1) BLANKET RECEIVING 100% COTION (1) CLAMP UMBILICAL CORD ST. (1) COVER TABLE REINFORCED 50" X 90" (1) SYRINGE 10CC W/0 NEEDLE L/LOCK (1) BOWL PLASTIC W/LID 80oz (2) GOWN XL SMS IMPERVIOUS REINFORCED (1) SHEET DRAPE 41" X 58" SMS (1) DRAPE C SECTION W/POUCH 102" X 121" X 78" (1) YANKAUER SUCTION TUBE W/0 VENT (1) WRAPPER 30" X 30" (15) SPONGE LAP PREWASH 18" X 18" XRD (1) TIME OUT BEACON NON WOVEN ST. (1) TUBE SUCTION CO NN ECT~" X 12' (2) LIGHT SHIELD (1) GOWN SURG REINF LARGE T/WRAP (2) PK. STRIP STERI CLOSURE W' X 4" (1) POLISHER CAUTERY TIP (1) DRESSING COMP OP 4" X 14" (1) COVER MAYO STAND REINFORCED (1) SKIN MARKER INK W/8 LABEL/TIME OUT/RULER EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents
- Recall number
- Z-2320-2014
- Recalling firm
- Customed, Inc
- Product
- CESAREAN PACK - - (7) TOWELS ABSORBENT 15" X 20" (2) SCALPEL WEIGHTED SAFETY #1 0 (1) BLANKET BABY 30" X 30" POLY ABS IMP (1) BOWL UTILITY QUART 32oz. (1) TUBE FEEDING 8Fr. 15" LONG (2) COUNTER NEEDLE/BLADE 10C MAG/CLEAR (1) BEANNIES BABY (2) PAD OBSTRETICAL XLGE ST. (1) SYRINGE 20cc W/0 NEEDLE LIL (1) SYRINGE EAR/ULCER 2oz. (1) BLANKET RECEIVING 100% COTION (1) CLAMP UMBILICAL CORD ST. (1) COVER TABLE REINFORCED 50" X 90" (1) SYRINGE 10CC W/0 NEEDLE L/LOCK (1) BOWL PLASTIC W/LID 80oz (2) GOWN XL SMS IMPERVIOUS REINFORCED (1) SHEET DRAPE 41" X 58" SMS (1) DRAPE C SECTION W/POUCH 102" X 121" X 78" (1) YANKAUER SUCTION TUBE W/0 VENT (1) WRAPPER 30" X 30" (15) SPONGE LAP PREWASH 18" X 18" XRD (1) TIME OUT BEACON NON WOVEN ST. (1) TUBE SUCTION CO NN ECT~" X 12' (2) LIGHT SHIELD (1) GOWN SURG REINF LARGE T/WRAP (2) PK. STRIP STERI CLOSURE W' X 4" (1) POLISHER CAUTERY TIP (1) DRESSING COMP OP 4" X 14" (1) COVER MAYO STAND REINFORCED (1) SKIN MARKER INK W/8 LABEL/TIME OUT/RULER EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents
- Status
- Terminated
- Initiated
- 2014-05-20
- Posted
- 2014-08-29
- Terminated
- 2017-08-14
- Reason
- Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
- Root cause
- Package design/selection
- Action
- Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
- Quantity
- 30 lots; 4396 units (multiple units per lot)
- Distribution
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot, serial or model codes
- Product code 900-403, 30 lots: 112020525 112030955 112031372 112051682 112062486 112093590 112104107 112125170 112125315 113015449 113015695 113025960 113036485 113046957 113057479 113057817 113068169 113078573 113088813 113099289 113099571 131110076 131110409 131110586 131210970 140211667 140311937 140312382 140412836 140513246
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OHM
- Firm FEI
- 2648727
- Firm city
- Fajardo
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28