DeVilbiss Healthcare LLC: DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter), 800 ML DISP Container, Unassembled, 800 ML DISP Container w/6' patient tubing, and 800 ML DISP Container w/6' patient tubing, assembled. DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter) is used to remove fluids from the airway or respiratory support system and infectious materials from wounds. It is available for use by consumers by physician order.
- Recall number
- Z-2319-2013
- Recalling firm
- DeVilbiss Healthcare LLC
- Product
- DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter), 800 ML DISP Container, Unassembled, 800 ML DISP Container w/6' patient tubing, and 800 ML DISP Container w/6' patient tubing, assembled. DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter) is used to remove fluids from the airway or respiratory support system and infectious materials from wounds. It is available for use by consumers by physician order.
- Status
- Terminated
- Initiated
- 2013-03-07
- Posted
- 2013-10-01
- Terminated
- 2014-11-06
- Reason
- End users of the DeVilbiss Disposable Suction Container/Filter had exposed the bacterial filter cartridge to fluid, thereby occluding the filter and rendering it unusable.
- Root cause
- Use error
- Action
- DeVilbiss Healthcare sent End users a field corrective action letter dated March 7, 2013, to all affected customers to reinforce the proper cleaning method. This letter is to inform you that we have come across cases where end users had exposed the bacterial filter cartridge in the disposable suction container to fluid, thereby occluding the filter and rendering it unusable Specifically what do you have to do when you receive a correction/recall notice? " The US FDA states in its correction/recall regulations that: "Consignees that receive a recall communication should immediately carry out the instructions set forth by the recalling firm and, where necessary, extend the recall to its consignees." "Enclosed you will find two Urgent Medical Device Correction letters addressed to DeVilbiss Dealer/Provider" and "DeVilbiss User" that we ask you read carefully. "It is important that you follow the actions outlined in these letters to ensure we can effectively disseminate this information to end-users. "To reach your direct accounts we ask that you send a copy of the "DeVilbiss User" letter to each of these end-users. We sincerely appreciate all of your efforts to ensure the safety of your customers that depend on DeVilbiss products. We thank you in advance for your assistance and apologize for any inconvenience that you may experience as a result of this situation. Should you have any questions or concerns, please contact DeVilbiss Healthcare Customer Service at 1-800-338-1988.
- Quantity
- 112,849
- Distribution
- Worldwide Distribution - USA and the countries of Mexico and Canada.
- Lot, serial or model codes
- Part Number 7305D-634 with model numbers 7305D-632, 7305D-633, and 7305D-633-25
- 510(k) numbers
- ["K982304"]
- PMA numbers
- not on file
- Product code
- JCX
- Firm FEI
- 2515872
- Firm city
- Somerset
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28