Skip to the content

FDA Medical Device Recalls (openFDA)

Mckesson Information Solutions LLC: McKesson Horizon Medical Imaging (HMI) Radiological Image Processing System, product versions 4.6.1 to 11.6., is a medical image and information management application intended to receive, transmit, store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from medical imaging systems.

Recall number
Z-2318-2013
Recalling firm
Mckesson Information Solutions LLC
Product
McKesson Horizon Medical Imaging (HMI) Radiological Image Processing System, product versions 4.6.1 to 11.6., is a medical image and information management application intended to receive, transmit, store, retrieve, display, print and process digital medical images, digital medical video, and associated medical information from medical imaging systems.
Status
Terminated
Initiated
2013-09-04
Posted
2013-09-30
Terminated
2017-02-06
Reason
Software malfunction may occur resulting in a certain combination of server configurations which has the potential to cause image loss.
Root cause
Software change control
Action
Consignees were notified via Urgent Field Safety Notice dated September 4. MMIC Customer support will be contacting customers known to have a multi-server configuration and will implement the configuration change.
Quantity
329 units
Distribution
Worldwide Distribution - United States including the territories of Guam and Puerto Rico and the countries of Canada and France.
Lot, serial or model codes
Products: HMI 4.6.1; HMI 5.x (5.0.1, 5.0.5, 5.0.6, 5.0.7, 5.0.8); HMI 11.0x (11.0.5, 11.0.6, 11.0.6 FR EP2, 11.0.7, 11.0.8); HMI 11.5x (11.5, 11.5.1, 11.5.2) and HMI 11.6.
510(k) numbers
["K043146"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3002684279
Firm city
Alpharetta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register