FDA Medical Device Recalls (openFDA)
GE Medical Systems, LLC: Allia IGS 5 Pulse angiographic X-ray system
- Recall number
- Z-2315-2026
- Recalling firm
- GE Medical Systems, LLC
- Product
- Allia IGS 5 Pulse angiographic X-ray system
- Status
- Open, Classified
- Initiated
- 2026-04-20
- Posted
- 2026-06-05
- Terminated
- not on file
- Reason
- GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
- Root cause
- Device Design
- Action
- An URGENT MEDICAL DEVICE CORRECTION dated 4/20/26 was sent to customers. Actions to be taken by Customer/User You can continue to use your system. As a reminder, qualified service personnel must follow the Service Manual applicable to your system and use the appropriate personal protective equipment (PPE) for electrical hazards when performing any activity inside the cabinet while power is ON. If any abnormal condition is observed on the Iris Power Cable during servicing, such as loose or visibly exposed wires, qualified service personnel must follow the Electrical Lockout/Tagout Procedure as described in Service Manual. Please ensure that all potential users and service personnel in your facility are made aware of this safety notification and the recommended actions and post this letter in a visible location near the C-FRT cabinet. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 18014 Digital Response Form or print, fill out manually, scan, and email to recall.18014@gehealthcare.com.
- Quantity
- 121 units
- Distribution
- Worldwide - US Nationwide distribution.
- Lot, serial or model codes
- UDI-DI: 00195278719263; Serial Numbers: M2-23-039, M2-23-056, M2-24-004, M2-24-019, M2-24-020, M2-24-025, M2-24-030, M2-24-031, M2-24-032, M2-24-033, M2-24-034, M2-24-037, M2-24-038, M2-24-039, M2-24-040, M2-24-046, M2-24-057, M2-24-062, M2-24-064, M2-24-066, M2-24-069, M2-24-071, M2-24-077, M2-24-078, M2-24-080, M2-24-083, M2-25-003, M2-25-004, M2-25-005, M2-25-011, M2-25-013, M2-25-015, M2-25-017, M2-25-019, M2-25-023, M2-25-024, M2-25-025, M2-25-026, M2-25-028, M2-25-032, M2-25-036, M2-25-037, M2-25-039, M2-25-040, M2-25-041, M2-25-042, M2-25-043, M2-25-046, M2-25-048, M2-25-049, M2-25-056, M2-25-057, M2-25-058, M2-25-062, M2-25-065, M2-25-067, M2-25-072, M2-25-074, M2-25-077, M2-25-078, M2-25-080, M2-25-250, M2-25-300, M2-26-002, M2-26-004, M2-26-006, M2-26-007, M2-UP2502, M3-22-012, M3-23-117, M3-24-019, M3-24-021, M3-24-023, M3-24-034, M3-24-052, M3-24-061, M3-24-074, M3-24-079, M3-24-086, M3-24-090, M3-24-094, M3-24-095, M3-24-097, M3-24-099, M3-24-101, M3-24-104, M3-24-111, M3-24-120, M3-24-128, M3-24-136, M3-24-137, M3-24-141, M3-24-143, M3-24-149, M3-24-151, M3-24-156, M3-24-159, M3-24-161, M3-24-163, M3-24-165, M3-24-168, M3-24-169, M3-24-172, M3-24-175, M3-25-003, M3-25-012, M3-25-017, M3-25-020, M3-25-027, M3-25-036, M3-25-061, M3-25-063, M3-25-069, M3-25-074, M3-25-076, M3-25-079, M3-25-080, M3-25-084, M3-25-085, M3-25-088, M3-25-089, M3-25-094, M3-25-104, M3-25-106, M3-25-107, M3-25-115, M3-25-118, M3-25-119, M3-25-120, M3-25-128, M3-25-130, M3-25-132, M3-25-145, M3-25-147, M3-25-148, M3-25-149, M3-25-150, M3-25-151, M3-26-002, M3-26-003, M3-26-005, M3-26-007, M3-26-021, M3-26-024, M3-26-027, M3-26-028, M3-26-033, M3-26-034, M3-UP2501, M3-UP2503
- 510(k) numbers
- ["K232344"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28