Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: Allia IGS 5 Pulse angiographic X-ray system

Recall number
Z-2315-2026
Recalling firm
GE Medical Systems, LLC
Product
Allia IGS 5 Pulse angiographic X-ray system
Status
Open, Classified
Initiated
2026-04-20
Posted
2026-06-05
Terminated
not on file
Reason
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Root cause
Device Design
Action
An URGENT MEDICAL DEVICE CORRECTION dated 4/20/26 was sent to customers. Actions to be taken by Customer/User You can continue to use your system. As a reminder, qualified service personnel must follow the Service Manual applicable to your system and use the appropriate personal protective equipment (PPE) for electrical hazards when performing any activity inside the cabinet while power is ON. If any abnormal condition is observed on the Iris Power Cable during servicing, such as loose or visibly exposed wires, qualified service personnel must follow the Electrical Lockout/Tagout Procedure as described in Service Manual. Please ensure that all potential users and service personnel in your facility are made aware of this safety notification and the recommended actions and post this letter in a visible location near the C-FRT cabinet. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 18014 Digital Response Form or print, fill out manually, scan, and email to recall.18014@gehealthcare.com.
Quantity
121 units
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 00195278719263; Serial Numbers: M2-23-039, M2-23-056, M2-24-004, M2-24-019, M2-24-020, M2-24-025, M2-24-030, M2-24-031, M2-24-032, M2-24-033, M2-24-034, M2-24-037, M2-24-038, M2-24-039, M2-24-040, M2-24-046, M2-24-057, M2-24-062, M2-24-064, M2-24-066, M2-24-069, M2-24-071, M2-24-077, M2-24-078, M2-24-080, M2-24-083, M2-25-003, M2-25-004, M2-25-005, M2-25-011, M2-25-013, M2-25-015, M2-25-017, M2-25-019, M2-25-023, M2-25-024, M2-25-025, M2-25-026, M2-25-028, M2-25-032, M2-25-036, M2-25-037, M2-25-039, M2-25-040, M2-25-041, M2-25-042, M2-25-043, M2-25-046, M2-25-048, M2-25-049, M2-25-056, M2-25-057, M2-25-058, M2-25-062, M2-25-065, M2-25-067, M2-25-072, M2-25-074, M2-25-077, M2-25-078, M2-25-080, M2-25-250, M2-25-300, M2-26-002, M2-26-004, M2-26-006, M2-26-007, M2-UP2502, M3-22-012, M3-23-117, M3-24-019, M3-24-021, M3-24-023, M3-24-034, M3-24-052, M3-24-061, M3-24-074, M3-24-079, M3-24-086, M3-24-090, M3-24-094, M3-24-095, M3-24-097, M3-24-099, M3-24-101, M3-24-104, M3-24-111, M3-24-120, M3-24-128, M3-24-136, M3-24-137, M3-24-141, M3-24-143, M3-24-149, M3-24-151, M3-24-156, M3-24-159, M3-24-161, M3-24-163, M3-24-165, M3-24-168, M3-24-169, M3-24-172, M3-24-175, M3-25-003, M3-25-012, M3-25-017, M3-25-020, M3-25-027, M3-25-036, M3-25-061, M3-25-063, M3-25-069, M3-25-074, M3-25-076, M3-25-079, M3-25-080, M3-25-084, M3-25-085, M3-25-088, M3-25-089, M3-25-094, M3-25-104, M3-25-106, M3-25-107, M3-25-115, M3-25-118, M3-25-119, M3-25-120, M3-25-128, M3-25-130, M3-25-132, M3-25-145, M3-25-147, M3-25-148, M3-25-149, M3-25-150, M3-25-151, M3-26-002, M3-26-003, M3-26-005, M3-26-007, M3-26-021, M3-26-024, M3-26-027, M3-26-028, M3-26-033, M3-26-034, M3-UP2501, M3-UP2503
510(k) numbers
["K232344"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register