Toshiba American Medical Systems Inc: Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.

Recall number
Z-2315-2018
Recalling firm
Toshiba American Medical Systems Inc
Product
Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Status
Terminated
Initiated
2018-01-04
Posted
not on file
Terminated
2023-10-10
Reason
The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Root cause
Software design
Action
Canon sent an Urgent - Medical Device Correction letter dated January 4, 2018. Customers were notified of the issue and informed that a modified firmware will be installed to correct the issue. When that modified firmware becomes available, the Toshiba service representative will contact customers to schedule an appointment to install the firmware. Until the modified firmware is installed, instructions were provided to confirm that the main power switch is sent to OFF, insert the power plug into the power outlet, and to set the main power switch to ON. The system should be operated according to the procedure described in section 9.1, Connection the Power Cable and Protective Grounding, in the operation manual. Additionally, customers were asked to complete and return the attached Customer Reply Form. For further questions, please call (714) 730-5000.
Quantity
162 units total
Distribution
USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.
Lot, serial or model codes
Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB1782054, HAB1732006, NJB1782062, NJB1782063, NJB1762004, NJB1762005, NJB1772022, JHA1752014, JHA1752016, JHA1752017, JHA1752018, HNA1792012, NJC17X2068, NJC17X2069, NJC17X2070, NJC17X2067, NJB1782046, NJB1782047, NJB1782048, NJB1782049, NJB1772039, NJB1782041, NJB1782045, NJC17X2083, NJC17X2084, NJC17X2081, NJC17X2079, NJC17X2080, NJC17X2082, NJC17X2076, NJC17X2077, NJC17X2078, NJB1772021, NJB1782044, JHA1752020, HNA1772005, NJB1782064, JHA1762032, HAB1732010, HAB1732008, NJB1782059, NJB1762008, NJB1762007, NJB1762009, NJB1772024, NJB1772014, NJB1772017, NJB1772018, NJB1772015, NJB1772016, NJB1772011, NJB1772012, NJB1772020, NJB1762010, NJB1772035, NJB1772036, NJB1772037, NJB1772038, NJB1772026, NJB1772027, NJB1772028, NJB1772030, NJB1772034, NJB1772025, NJB1772032, NJB1772013, JHA1762036, NJC17X2090, JHA1762033, HNA1772010, HNA1772009, NJB1772029, HNA1772004, HNA1772003, NJB1782065, NJC17X2071, NJB17822040, NJC17X2074, NJC17X2075, NJC17X2085, HAB1732026, JHA16X2007, HAB1732022, JHA1752023, JHA1762026, JHA1762025, HAB1732028, JHA16X2001, JHA16X2003, HAB1742011, HAB1742012, JHA16X2001, JHA16X2003, JHA1752004, JHA1752005, JHA1752006, JHA1752007, JHA1752008, JHA1752009, JHA1752010, JHA1752011, JHA1752012, JHA1752013, NJB1772033, NJC17X2072, NJC17X2073, NJC17X2088, HAB16Z2001, HAB16Z2004, HAB16Z2005, HAB1732024, HAB1732032, HNA1792015, HNB17X2016, HNB17X2017, HNB17X2018, HNB17X2019, HNB17X2020, HNB17X2021, HNB17X2022, HNB17X2023, HNB17X2024, HNB17X2025, HNB17X2026, HNB17X2027, HNB17X2028, HNB17X2029, HNB17X2030, JHA16X2002, JHA16X2005, JHA16X2006, JHA16X2008, JHA16X2009, JHA1752011, JHA1752012, JHA1752013, JHA1752015, JHA1762027, JHA1762028, JHA1762035, JHA1762037, JHA1762038, JHA1762039, HAB1742001, HAB1742002, JHA16X2001
510(k) numbers
["K163702"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register