Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.

Recall number
Z-2315-2012
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.
Status
Terminated
Initiated
2012-02-21
Posted
2012-09-06
Terminated
2012-09-07
Reason
The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.
Root cause
Component change control
Action
Toshiba America Medical Systems, Inc, sent a Urgent Medical Device Correction letter dated February 21, 2012, via USPS return receipt mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed if you subsequently find that images are abnormally noisy compared to images acquired under the same conditions in the system before this modification was performed, contact your Toshiba Service Representative. Please share this information with all users and reviewing radiologist as well as clinical engineering or biomedical group at your facility. Please complete and return the attached form and fax it to the toll free number at the top of the form. This form can also be sent via email toraffairs@tams.com. Contact information was also provided (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.
Quantity
3 units
Distribution
US (nationwide) distribution including the states of Ohio and Texas.
Lot, serial or model codes
MRT-1503/P5-Vantage MRI System.  Serial Codes: P5A0632002, P5B06Y2005, P5B0712006, P3F11Z2143.
510(k) numbers
["K060003"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register