Skip to the content

FDA Medical Device Recalls (openFDA)

AdvanDx, Inc.: E.faecalis/OE PNA FISH Intended Use: E. faecalis/OE PNA FISH" (Cat. No. KT003) is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Enterococcus faecalis and selected other enterococci (OE) from blood cultures. E. faecalis/OE PNA FISH provides rapid identification of E. faecalis and selected other enterococci (OE) on smears made from positive blood cultures and is compatible with all major blood culture media

Recall number
Z-2315-2010
Recalling firm
AdvanDx, Inc.
Product
E.faecalis/OE PNA FISH Intended Use: E. faecalis/OE PNA FISH" (Cat. No. KT003) is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Enterococcus faecalis and selected other enterococci (OE) from blood cultures. E. faecalis/OE PNA FISH provides rapid identification of E. faecalis and selected other enterococci (OE) on smears made from positive blood cultures and is compatible with all major blood culture media
Status
Terminated
Initiated
2010-06-08
Posted
2010-08-31
Terminated
2011-06-06
Reason
Some strains of streptococus anginosus cross-reach with E. faecalis/OE PNA FISH to produce a false positive green signal.
Root cause
Device Design
Action
The firm, AdvanDx, sent a letter dated June 8, 2010, and email notification on June 11, 2010, to all customers. The letter described the product, problem and action to be taken by customers. AdvanDx notified the customers that there was a change to the package insert under "limitations" to include the statement "some strains of Streptococcus anginosus produce a false green positive fluorescence due to sequence similarities and enclosed a package insert for all E. facealis/OE PNA FISH (KT003) kits. The customers were instructed to replace the Package Insert (revision B) included with any kits in their inventory with the enclosed Package Insert (revision C). Note: All shipments of E. faecalis/OE PNA FISH, as of June 14, 2010, will contain this new package insert. If you have any questions or requests for further information, please contact AdvanDx Technical Support at US (Toll Free): 1-800-376-0009 or email techsupport@advandx.com.
Quantity
541 units
Distribution
Worldwide distribution: USA and countries including Danmark, Finland and Sverige (European).
Lot, serial or model codes
Catalog number KT003
510(k) numbers
["K083074"]
PMA numbers
not on file
Product code
OAH
Firm FEI
3004080598
Firm city
Woburn
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register