Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: Plum Duo Precision IV Pump, 40002-0403

Recall number
Z-2313-2026
Recalling firm
ICU Medical, Inc.
Product
Plum Duo Precision IV Pump, 40002-0403
Status
Open, Classified
Initiated
2026-04-29
Posted
2026-06-05
Terminated
not on file
Reason
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Root cause
Software Design Change
Action
On April 29, 2026, the firm notified consignees via letters titled "URGEND MEDICAL DEVICE CORRECTION." Customers were instructed to identify all affected pumps in their possession and ensure all users or potential users of the pumps are immediately made aware of the notification. During bolus programming, clinicians should check whether the soft limit override icon is displayed next to the bolus dose entry field on the programming screen. If the pump displays the icon, the clinician should review the program and make any necessary adjustments before confirming the program. The firm is developing a software update to correct the issue and will notify customers when it is available.
Quantity
13,613
Distribution
Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
Lot, serial or model codes
UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3
510(k) numbers
["K242115"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register