Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Molecular: VP 2000 Processor Heated Reagent Basins; automated slide stainer; Abbott Molecular Inc., Des Plaines, IL 60018; The basins are sold under the the following part numbers: a) list 02J11-013, part number 30-144152 (replacement basins); b) list 02J11-060, part number 30-144100 (as a component of the 100 volt VP 2000 Processor); c) list 02J11-001, part number 30-144101 (as a component of the 117 volt VP 2000 Processor); d) list 02J11-004, part number 30-144102 (as a component of the 230 volt VP 2000 Processor); Designed to automate and standardize FISH slide specimen processing and routine slide staining for the laboratory.

Recall number
Z-2313-2010
Recalling firm
Abbott Molecular
Product
VP 2000 Processor Heated Reagent Basins; automated slide stainer; Abbott Molecular Inc., Des Plaines, IL 60018; The basins are sold under the the following part numbers: a) list 02J11-013, part number 30-144152 (replacement basins); b) list 02J11-060, part number 30-144100 (as a component of the 100 volt VP 2000 Processor); c) list 02J11-001, part number 30-144101 (as a component of the 117 volt VP 2000 Processor); d) list 02J11-004, part number 30-144102 (as a component of the 230 volt VP 2000 Processor); Designed to automate and standardize FISH slide specimen processing and routine slide staining for the laboratory.
Status
Terminated
Initiated
2010-07-20
Posted
2010-08-31
Terminated
2012-04-19
Reason
The VP2000 Processor Heated Reagent Basins crack along the bottom seam, have non-uniform coating of interior corners, may be warped, and some may have incorrect dimensions.
Root cause
Reprocessing Controls
Action
Abbott sent Urgent Product Recall letters dated 7/20/10 to the VP 2000 Processor customers on same date via FedEx, informing them that the heated reagent basins could crack along the bottom seam, have non-uniform coating of interior corners, warping of the aluminum housing or incorrect dimensions. These deficiencies could result in leakage or the inability to place the basins into the VP 2000 Processor, resulting in possible exposure to chemical reagents, a slippery floor hazard, and/or delay or failure to start a protocol run. An Abbott Molecular customer service representative will call them to coordinate replacement of all of their heated reagent basins before 10/31/10. Customers were to complete and return the Customer Reply form provided. Any questions were directed to their local Abbott Molecular Representative. *** A second recall letter was sent to all customers on 12/21/10 explaining that the VP 2000 heated reagent basins previously inspected and determined to be acceptable could develop a leak, and that they would be contacted by an Abbott Molecular Customer Service Representative or Field Service representative in early 2011 to coordinate replacement of all of their VP 2000 heated reagent basins. ***
Quantity
219 basins
Distribution
Worldwide Distribution -- Australia, Belgium, France, Germany, India, Italy, Japan, Pakistan, Poland, Russia, Singapore, South Korea, Switzerland, Taiwan, Thailand, United Kingdom, Uruguay, and the United States.
Lot, serial or model codes
All basins distributed since September 2008.
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
3005248192
Firm city
Des Plaines
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register