Skip to the content

FDA Medical Device Recalls (openFDA)

St. Jude Medical Cardiovascular Division: The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot 2056960, Manufacturer St. Jude Medical 14901 DeVeau Place, Minnetonka, MN 55345-2126 USA, Sterile EO The St. Jude Medical Livewire TC" Ablation Catheter is a flexible electrode catheter constructed of a radiopaque polyurethane insulation/shaft and incorporated platinum electrode. The distal tip portion may be deflected by a remote control handle located a t the proximal end of t the catheter.

Recall number
Z-2313-2008
Recalling firm
St. Jude Medical Cardiovascular Division
Product
The St. Jude Medical Livewire TC" Ablation Catheter, 8mm Tip w/Thermocouple & Thermistor REF 402196, Length 115 cm, Lot 2056960, Manufacturer St. Jude Medical 14901 DeVeau Place, Minnetonka, MN 55345-2126 USA, Sterile EO The St. Jude Medical Livewire TC" Ablation Catheter is a flexible electrode catheter constructed of a radiopaque polyurethane insulation/shaft and incorporated platinum electrode. The distal tip portion may be deflected by a remote control handle located a t the proximal end of t the catheter.
Status
Terminated
Initiated
2008-06-05
Posted
2008-09-20
Terminated
2012-02-04
Reason
St. Jude Medical has initiated a recall of one lot of Livewire TC" Ablation Catheter. The product was mis-labelled as having an 8 mm tip instead of 4 mm tip. This could potentially result in elevated temperatures and formation of coagulum at the tip electrode. If this coagulum were to detach from the product, the worst case this could potentially result in embolization.
Root cause
Labeling mix-ups
Action
Consignees were sent a "St Jude Medical Urgent Product Recall" letter addressed to "Dear Doctor" and a "Doctor's Acknowledgement Form" . The letter described the product, problem and the possibility of a health hazard. They also requested discontinue of product and gave instructions for return of product. Contact St. Jude Medical at 952-933-4700 if you have questions.
Quantity
39
Distribution
Canada and Belgium
Lot, serial or model codes
Lot # 2056960
510(k) numbers
not on file
PMA numbers
["P960016"]
Product code
LPB
Firm FEI
2182269
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register