Skip to the content

FDA Medical Device Recalls (openFDA)

OPTI Medical Systems, Inc: OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Recall number
Z-2310-2021
Recalling firm
OPTI Medical Systems, Inc
Product
OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004
Status
Open, Classified
Initiated
2021-06-25
Posted
not on file
Terminated
not on file
Reason
The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.
Root cause
Nonconforming Material/Component
Action
/The firm disseminated a technical bulletin beginning on 06/25/2021 by email identifying the /problem and providing additional instructions to be implemented when using the product.
Quantity
3552 kits
Distribution
Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados
Lot, serial or model codes
a) 99-57003, Lot Numbers: 20808, 20812, 20813 b) 99-57004, Lot Numbers: 20881, 20901, 211013
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
3004102403
Firm city
Roswell
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register