Skip to the content

FDA Medical Device Recalls (openFDA)

SSC- Surgical Specialties Corporation: Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife

Recall number
Z-2309-2026
Recalling firm
SSC- Surgical Specialties Corporation
Product
Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife
Status
Open, Classified
Initiated
2026-04-23
Posted
2026-06-04
Terminated
not on file
Reason
3.0mm surgical knives labeled and packaged as 2.2mm surgical knives.
Root cause
Mixed-up of materials/components
Action
On August 22, 2026, Surgical Specialties Corp. (subsidiary of Corza Medical) issued a Urgent Medical Device recall notification via E.Mail to affected consignees. Surgical Specialties ask consignees to take the following actions: 1. Immediately examine your inventory and quarantine product subject to recall. We request the return of any affected product that may still be in your inventory. We will replace the product 2. Recall Acknowledgement & Receipt Please complete and return the attached Medical Device Recall Acknowledgement & Receipt form as soon as possible to returns.corzaeye@corza.com. 3. To initiate a return, contact returns.corzaeye@corza.com to process the request. 4. If you have any questions related to product quality, contact us at complaints@corza.com.
Quantity
1818 units
Distribution
Worldwide - US Nationwide distribution in the states of Texas, Utah, Michigan and the countries of Turkey, Germany, Romania, Poland, United Kingdom.
Lot, serial or model codes
UDI-DI 00848782027514; Lot Number FA22CHQ
510(k) numbers
not on file
PMA numbers
not on file
Product code
HNN
Firm FEI
3011767135
Firm city
Chula Vista
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register