GE Medical Systems China Co., Ltd.: MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph

Recall number
Z-2306-2026
Recalling firm
GE Medical Systems China Co., Ltd.
Product
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
Status
Open, Classified
Initiated
2026-04-21
Posted
2026-06-03
Terminated
not on file
Reason
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
Root cause
Software Manufacturing/Software Deployment
Action
Firm notified consignees beginning April 21, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION". Customers were given instructions on how to check their software version. If a customer's software version begins with 3.00, the customer should not use the system until a GE HealthCare representative has updated the device with the latest licensed software version.
Quantity
16
Distribution
Worldwide - US Nationwide distribution in the states of Florida and Georgia. The countries of India, Italy, Republic of Korea, and United Kingdom.
Lot, serial or model codes
Software version 3.00 855001-001, UDI-DI 00195278276070 8855002-001, UDI-DI 00195278276124
510(k) numbers
["K221321"]
PMA numbers
not on file
Product code
DXH
Firm FEI
3003513129
Firm city
Wuxi
Firm state
not on file
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register