Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, Wisconsin. Merge PACS 6.0 is software intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.

Recall number
Z-2303-2016
Recalling firm
Merge Healthcare, Inc.
Product
Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, Wisconsin. Merge PACS 6.0 is software intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.
Status
Terminated
Initiated
2016-01-30
Posted
2016-07-29
Terminated
2021-05-07
Reason
A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
Root cause
Software design
Action
The recalling firm, Merge, issued an "URGENT: MEDICAL DEVICE RECALL" letter dated 1/27/2016 via e-mail on 1/30/2016 or via certified mail if they did not have an e-mail address for the customers. The letter described the product, problem and actions to be taken. The customers were instructed to not discontinue use of Merge PACS; to open prior comparison studies in the secondary viewer to check if there were annotations and read the report or visit note pertaining to that study to see if there are noted findings that were not otherwise apparent; to receive an upgrade call Merge Customer Service at (877) 741-5369 or send an email to support@merge.com to schedule the upgrade (Reference this call #2016-010); complete the online form: http://www.merge.com/Landing-Pages/Customer-Upgrade-Decline-Form.aspx no later than 2/22/2016; If you decide to decline the maintenance release, complete the attached Customer Decline Response form and return to Merge Healthcare via recall@merge.com, and ensure all users of the product are provided with this notification. If you have further distributed this product, identify your customers and notify them at once of this product recall. If you have any additional questions, call Merge Customer Service at (877) 741-5369 or support@merge.com.
Quantity
208 sites have the affected versions
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Australia, Belgium, Canada, Jordan, New Zealand, and the United Kingdom.
Lot, serial or model codes
Versions: V6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9, 6.6.0, 6.6.1, 6.6.1.1, 6.6.2
510(k) numbers
["K082144"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register