FDA Medical Device Recalls (openFDA)
Philips Healthcare Inc.: Philips SureSigns VSi/VS2+ patient monitors Model Part Numbers VSi/ 863275, 863276, 863277, VS2+ 863278, 863279. The SureSigns VS2+ vital signs monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility. The SureSigns VSi vital signs monitor is for measurement of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility
- Recall number
- Z-2303-2012
- Recalling firm
- Philips Healthcare Inc.
- Product
- Philips SureSigns VSi/VS2+ patient monitors Model Part Numbers VSi/ 863275, 863276, 863277, VS2+ 863278, 863279. The SureSigns VS2+ vital signs monitor is for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility. The SureSigns VSi vital signs monitor is for measurement of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility
- Status
- Terminated
- Initiated
- 2012-08-15
- Posted
- 2012-08-30
- Terminated
- 2014-03-13
- Reason
- Philips SureSigns VSi/VS2+ patient monitors may not be properly secured and may fall from the wall mount
- Root cause
- Process design
- Action
- Philips sent an "URGENT- MEDICAL DEVICE CORRECTION" letter dated August 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers using the VSi/VS2+ with the GCX wall mount are instructed to complete and fax a Customer Reply Wall Mount Order Card to the Business Unit or Key Market indicating the number of wall mounts in use. Upon receipt of the Customer Reply Wall Mount Order Card, Philips will send a replacement of the mounting hardware. Customers should contact their local Philips representative for questions regarding this recall.
- Quantity
- 2314 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, DENMARK, EL SALVADOR, FRANCE, GERMANY, ICELAND, INDIA, IRELAND ISRAEL, ITALY, JORDAN, KENYA, LEBANON, MALAYSIA,NAMIBIA NETHERLANDS, NORWAY, POLAND, PORTUGAL, RWANDA SAUDI ARABIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN SWITZERLAND, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Lot, serial or model codes
- Serial Number Range CN14800101-CN14801360 CN21201362-CN21202440, CN21202525, SP13701043, USPP101007
- 510(k) numbers
- ["K112652"]
- PMA numbers
- not on file
- Product code
- DXN
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28