Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, Lp: 3DKnee e+ Tibial Insert - HXLPE-VE, Sterile, Material: Highly cross linked vitamin E PE, Sz 2/9mm. The insert is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along lateral condyle. This insert is to be used along with the other components that comprise the 3DKnee System.

Recall number
Z-2302-2012
Recalling firm
Encore Medical, Lp
Product
3DKnee e+ Tibial Insert - HXLPE-VE, Sterile, Material: Highly cross linked vitamin E PE, Sz 2/9mm. The insert is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along lateral condyle. This insert is to be used along with the other components that comprise the 3DKnee System.
Status
Terminated
Initiated
2012-08-09
Posted
2012-08-30
Terminated
2013-01-25
Reason
The firm is recalling one lot of 392-09-706, 3DKnee e+ Tibial insert, 9 mm, Size 6, Right. The shelf life labeling of specific lots for a component were found to be incorrect and were found to be expired.
Root cause
Incorrect or no expiration date
Action
DJO Global sent an Urgent Field Safety Notice dated August 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the Notice to all those who need to be aware within their organization or to any organization where the potentially affected product has been transferred. Customers were instructed to contact the firm to place a replacement order and receive an RMA number. For questions customers were asked to call 512-834-6255. For questions regarding this recall call 512-832-9500.
Quantity
7 units
Distribution
Worldwide Distribution including Venezuela.
Lot, serial or model codes
Model Number 392-09-706, Lot Number 59602519
510(k) numbers
["K091956"]
PMA numbers
not on file
Product code
OIY
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register