FDA Medical Device Recalls (openFDA)

Hologic, Inc.: Selenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer. The Selenia Full Field Digital Mammography System is intended for use in the same clinical applications as traditional screen-film mammography systems. Mammography images can be interpreted on either hard copy film or soft copy review workstations, cleared for mammography

Recall number
Z-2301-2013
Recalling firm
Hologic, Inc.
Product
Selenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer. The Selenia Full Field Digital Mammography System is intended for use in the same clinical applications as traditional screen-film mammography systems. Mammography images can be interpreted on either hard copy film or soft copy review workstations, cleared for mammography
Status
Terminated
Initiated
2013-09-06
Posted
2013-09-30
Terminated
2013-12-13
Reason
Software: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most recently dated images
Root cause
Software design
Action
Hologic sent an Urgent Medical Device Recall letter dated September 5, 2013, by Certified Mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised how to avoid loss of images until Hologic corrects the problem. A software upgrade to customers who are connected to Hologic Connect who have already been connected remotely and customers who have already been connected via a Service Call will receive notification of the problem and the date the system was remotely corrected notified by Certified Mail. A Dealer notification issued by certified mail to inform Dealers of the need to notify their customers of the problem and to correct the identified Selenia system using Hologic Technical Bulletin TB-00072. Selenia software version 5.2 will implement a double safeguard to both guarantee a 28 day retention period, as well as ensure images are removed in a least recently used order. If there are any questions regarding this field correction, please call Hologic Customer Support at 877 371-4372.
Quantity
62 systems
Distribution
Worldwide Distribution - USA including the states of AL, AR, CA, CO, FL, IL, KY, LA, NE, NC, NJ, NY, OH, PA, TN, TX, UT, WA and WV., and the countries of Canada, Denmark, Iran, Italy, Mexico and Taiwan.
Lot, serial or model codes
Software versions 5.0.x and 5.1.x
510(k) numbers
not on file
PMA numbers
["P010025"]
Product code
MUE
Firm FEI
1220984
Firm city
Danbury
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register