FDA Medical Device Recalls (openFDA)
Ge Healthcare It: Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.
- Recall number
- Z-2300-2016
- Recalling firm
- Ge Healthcare It
- Product
- Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.
- Status
- Terminated
- Initiated
- 2013-02-18
- Posted
- not on file
- Terminated
- 2017-12-13
- Reason
- A software defect was discovered that causes images to be out of context with clinical information.
- Root cause
- Software design
- Action
- GE sent an "Urgent Medical Device Correction" letter to all affected customers on April 2, 2013.The letter identified the product the problem and the action customers need to take. The software has been modified to correct these defects. The modified software will be released and installed in user sites through GE Healthcare FMI #85204. For further questions, please call ( 847) 277-4468.
- Quantity
- 13
- Distribution
- US Nationwide Distribution.
- Lot, serial or model codes
- Imagecast PACS 3.5.69, 3.5.86, 3.5.87 and 3.6 with Centricity. RIS-IC 10.6 versions prior to update Package 18.
- 510(k) numbers
- ["K052618"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 3004526608
- Firm city
- Barrington
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28