Skip to the content

FDA Medical Device Recalls (openFDA)

Hardy Diagnostics: HardyCHROM MRSA,REF; G307, LOT: 447694, A selective and differential chromogenic medium for the qualitative direct detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA). Not intended to diagnose infection of guide therapy. Further testing required for susceptibility or typing. Container Type: 15 x 100mm monoplate Packaged: 10 plates/sleeve STORAGE: 2-8 degree C on receipt, LIGHT SENSITIVE IVD UDI: (01)08165760201400 (10)447694 (17)191206

Recall number
Z-2298-2020
Recalling firm
Hardy Diagnostics
Product
HardyCHROM MRSA,REF; G307, LOT: 447694, A selective and differential chromogenic medium for the qualitative direct detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA). Not intended to diagnose infection of guide therapy. Further testing required for susceptibility or typing. Container Type: 15 x 100mm monoplate Packaged: 10 plates/sleeve STORAGE: 2-8 degree C on receipt, LIGHT SENSITIVE IVD UDI: (01)08165760201400 (10)447694 (17)191206
Status
Terminated
Initiated
2019-11-14
Posted
2020-06-09
Terminated
2021-04-28
Reason
False Positive results due to the MSSA (methicillin-susceptible) strain Staphylococcus aureus (ATCC 29213) breakthrough when this strain should have been inhibited.
Root cause
Nonconforming Material/Component
Action
On 11/14/2019, the firm sent a notification of recall letter to customers, informing them of the performance issue of the device lot of the breakthrough on firm's differential chromogenic medium. The customers were advised to discard the remaining stock, and notify the Technical Services Department. If the customer would like replacement or credit, contact Technical Services Department or via e-mail at TechService@HardyDiagnostics.com, or 800-266-2222 option 2 or ext. 5658.
Quantity
3,580 ea (358 kits, pack of 10 in each)
Distribution
U.S. Distribution in the states of WI, CA, WA, IA, MT, UT, AZ, WY, KS, MO, OR, TX, VI, including PR and Guam. OUS: British Virgin Islands, Kenya,
Lot, serial or model codes
Catalogue no.: G307, Lot number 447694 HDx, UDI # 00816576020140
510(k) numbers
["K102922"]
PMA numbers
not on file
Product code
JSO
Firm FEI
2022807
Firm city
Santa Maria
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register