FDA Medical Device Recalls (openFDA)
Merit Medical Systems, Inc.: Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029
- Recall number
- Z-2294-2025
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029
- Status
- Open, Classified
- Initiated
- 2025-06-05
- Posted
- 2025-08-12
- Terminated
- not on file
- Reason
- High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
- Root cause
- Process control
- Action
- On 5/5/2025, recall notices were emailed to customers who were asked to do the following: 1) Identify affected devices and quarantine, discontinue use and distribution. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4. Return all affected devices to the firm. 5. Complete, and return the customer response form via email to response@merit.com If you have questions, contact the firm's customer service at RESPONSE@merit.com or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri
- Quantity
- 2779
- Distribution
- US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.
- Lot, serial or model codes
- REF/UDI-DI/Lot(Expiration): K04-00417/00884450046630/I3114667(12/10/2027), I3147550(1/28/2028); K04-00963/00884450046760/H3117053(12/14/2027), H3138590(1/20/2028), H3154108(2/10/2028), H3154786(2/10/2028), H3162244(2/19/2028), H3166789(3/2/2028), H3173526(3/23/2028); K04-01029/00884450148938/H3117059(12/17/2027), H3154112(2/10/2028), H3159017(2/19/2028), H3166777(3/9/2028), H3181994(4/1/2028);
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJA
- Firm FEI
- 1721504
- Firm city
- South Jordan
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28